A glucose electrode using high-stability glucose-oxidase collagen membranes - École des Ponts ParisTech Accéder directement au contenu
Article Dans Une Revue Diabetes Care Année : 1982

A glucose electrode using high-stability glucose-oxidase collagen membranes

Résumé

A very sensitive glucose electrode was developed using a glucose-oxidase membrane and an anodically polarized platinum disk. Calibration curves were linear over 4.5 concentration decades. It was adapted for human blood samples. The lifetime of glucose-oxidase collagen membranes was greater than 3 yr at 4°C and was ~ 6 month at 20-30°C.
Fichier principal
Vignette du fichier
Thevenot-Glucose-electrode-Diabetes-Care-1982-5-203-206.pdf (460.49 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

hal-01083014 , version 1 (14-11-2014)

Licence

Paternité

Identifiants

Citer

Daniel R. Thevenot, Robert Sternberg, Pierre Coulet. A glucose electrode using high-stability glucose-oxidase collagen membranes. Diabetes Care, 1982, 5 (3), pp.203 - 206. ⟨10.2337/diacare.5.3.203⟩. ⟨hal-01083014⟩