Abstract : A very sensitive glucose electrode was developed using a glucose-oxidase membrane and an anodically polarized platinum disk. Calibration curves were linear over 4.5 concentration decades. It was adapted for human blood samples. The lifetime of glucose-oxidase collagen membranes was greater than 3 yr at 4°C and was ~ 6 month at 20-30°C.
https://hal-enpc.archives-ouvertes.fr/hal-01083014 Contributeur : Daniel ThevenotConnectez-vous pour contacter le contributeur Soumis le : vendredi 14 novembre 2014 - 18:30:40 Dernière modification le : samedi 15 janvier 2022 - 04:02:09 Archivage à long terme le : : dimanche 15 février 2015 - 11:35:29
Daniel Thevenot, Robert Sternberg, Pierre Coulet. A glucose electrode using high-stability glucose-oxidase collagen membranes. Diabetes Care, American Diabetes Association, 1982, 5 (3), pp.203 - 206. ⟨10.2337/diacare.5.3.203⟩. ⟨hal-01083014⟩